tables:
  - file: "max31856/tables/table-00-p2-abs_max.csv"
    title: "Absolute Maximum Ratings"
    page: 2
  - file: "max31856/tables/table-03-p2-spec.csv"
    title: "Electrical Characteristics"
    page: 2
  - file: "max31856/tables/table-04-p10-pin.csv"
    title: "Pin Description"
    page: 10
  - file: "max31856/tables/table-05-p12-data.csv"
    title: "Table 1. Supported Thermocouples and Temperature Ranges"
    page: 12
  - file: "max31856/tables/table-06-p13-data.csv"
    title: "Table 2. Reference Junction (Cold-Junction) Temperature Data Format"
    page: 13
  - file: "max31856/tables/table-07-p13-data.csv"
    title: "Table 3. Linearized Thermocouple Temperature Data Format"
    page: 13
  - file: "max31856/tables/table-08-p14-data.csv"
    title: "Table 4. Open-Circuit Detection Mode"
    page: 14
  - file: "max31856/tables/table-09-p15-data.csv"
    title: "Table 5. Serial Interface Function"
    page: 15
  - file: "max31856/tables/table-10-p18-register.csv"
    title: "Table 6. Register Memory Map"
    page: 18
  - file: "max31856/tables/table-11-p18-register.csv"
    title: "Register 00h/80h: Configuration 0 Register (CR0) bit map"
    page: 18
  - file: "max31856/tables/table-12-p19-register.csv"
    title: "Register 00h/80h: Configuration 0 Register (CR0) (continued) — Bit Descriptions"
    page: 19
  - file: "max31856/tables/table-13-p20-register.csv"
    title: "Register 01h/81h: Configuration 1 Register (CR1) bit map"
    page: 20
  - file: "max31856/tables/table-14-p20-register.csv"
    title: "Register 01h/81h: Configuration 1 Register (CR1) — Bit Descriptions"
    page: 20
  - file: "max31856/tables/table-15-p21-register.csv"
    title: "Register 02h/82h: Fault Mask Register (MASK) bit map"
    page: 21
  - file: "max31856/tables/table-16-p21-register.csv"
    title: "Register 02h/82h: Fault Mask Register (MASK) — Bit Descriptions"
    page: 21
  - file: "max31856/tables/table-17-p22-register.csv"
    title: "Register 03h/83h: Cold-Junction High Fault Threshold Register (CJHF) bit map"
    page: 22
  - file: "max31856/tables/table-18-p22-register.csv"
    title: "Register 04h/84h: Cold-Junction Low Fault Threshold Register (CJLF) bit map"
    page: 22
  - file: "max31856/tables/table-19-p22-register.csv"
    title: "Register 05h/85h: Linearized Temperature High Fault Threshold Register, MSB (LTHFTH) bit map"
    page: 22
  - file: "max31856/tables/table-20-p22-register.csv"
    title: "Register 06h/86h: Linearized Temperature High Fault Threshold Register, LSB (LTHFTL) bit map"
    page: 22
  - file: "max31856/tables/table-21-p23-register.csv"
    title: "Register 07h/87h: Linearized Temperature Low Fault Threshold Register, MSB (LTLFTH) bit map"
    page: 23
  - file: "max31856/tables/table-22-p23-register.csv"
    title: "Register 08h/88h: Linearized Temperature Low Fault Threshold Register, LSB (LTLFTL) bit map"
    page: 23
  - file: "max31856/tables/table-23-p23-register.csv"
    title: "Register 09h/89h: Cold-Junction Temperature Offset Register (CJTO) bit map"
    page: 23
  - file: "max31856/tables/table-24-p24-register.csv"
    title: "Register 0Ah/8Ah: Cold-Junction Temperature Register, MSB (CJTH) bit map"
    page: 24
  - file: "max31856/tables/table-25-p24-register.csv"
    title: "Register 0Bh/8Bh: Cold-Junction Temperature Register, LSB (CJTL) bit map"
    page: 24
  - file: "max31856/tables/table-26-p24-register.csv"
    title: "Register 0Ch: Linearized TC Temperature, Byte 2 (LTCBH) bit map"
    page: 24
  - file: "max31856/tables/table-28-p25-register.csv"
    title: "Register 0Eh: Linearized TC Temperature, Byte 0 (LTCBL) bit map"
    page: 25
  - file: "max31856/tables/table-29-p25-register.csv"
    title: "Register 0Fh: Fault Status Register (SR) bit map"
    page: 25
  - file: "max31856/tables/table-30-p26-register.csv"
    title: "Register 0Fh: Fault Status Register (SR) (continued) — Bit Descriptions"
    page: 26
  - file: "max31856/tables/table-31-p29-data.csv"
    title: "Ordering Information"
    page: 29
  - file: "max31856/tables/table-32-p29-data.csv"
    title: "Package Information"
    page: 29
  - file: "max31856/tables/table-33-p30-data.csv"
    title: "Revision History"
    page: 30
  - file: "max31856/tables/table-01-p2-data.csv"
    title: "Package Thermal Characteristics (Note 1)"
    page: 2
  - file: "max31856/tables/table-02-p2-spec.csv"
    title: "Recommended DC Operating Conditions"
    page: 2
  - file: "max31856/tables/table-27-p25-register.csv"
    title: "Register 0Dh: Linearized TC Temperature, Byte 1 (LTCBM) bit map"
    page: 25
figures:
  - file: "max31856/figures/max31856-f000.png"
    caption: ""
    type: "application_circuit"
    page: 1
  - file: "max31856/figures/max31856-f004.png"
    caption: "Figure 1. Timing Diagram: SPI Read Data Transfer"
    type: "timing_diagram"
    page: 6
  - file: "max31856/figures/max31856-f005.png"
    caption: "Figure 2. Timing Diagram: SPI Write Data Transfer"
    type: "timing_diagram"
    page: 6
  - file: "max31856/figures/max31856-f006.png"
    caption: "ACTIVE SUPPLY CURRENT vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 7
  - file: "max31856/figures/max31856-f007.png"
    caption: "STANDBY SUPPLY CURRENT vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 7
  - file: "max31856/figures/max31856-f008.png"
    caption: "THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc04)"
    type: "characteristic_curve"
    page: 7
  - file: "max31856/figures/max31856-f009.png"
    caption: "AV = 8 FULL-SCALE ERROR vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 7
  - file: "max31856/figures/max31856-f010.png"
    caption: "THERMOCOUPLE INPUT BIAS CURRENT vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 7
  - file: "max31856/figures/max31856-f011.png"
    caption: "THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc05)"
    type: "characteristic_curve"
    page: 7
  - file: "max31856/figures/max31856-f012.png"
    caption: "AV = 32 FULL-SCALE ERROR vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 7
  - file: "max31856/figures/max31856-f013.png"
    caption: "AV = 8 OFFSET ERROR vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 8
  - file: "max31856/figures/max31856-f014.png"
    caption: "FULL-SCALE ERROR vs. COMMON-MODE VOLTAGE"
    type: "characteristic_curve"
    page: 8
  - file: "max31856/figures/max31856-f015.png"
    caption: "B-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 8
  - file: "max31856/figures/max31856-f016.png"
    caption: "AV = 32 OFFSET ERROR vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 8
  - file: "max31856/figures/max31856-f017.png"
    caption: "COLD-JUNCTION TEMPERATURE ERROR vs. TEMPERATURE"
    type: "characteristic_curve"
    page: 8
  - file: "max31856/figures/max31856-f018.png"
    caption: "E-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 8
  - file: "max31856/figures/max31856-f019.png"
    caption: "J-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 9
  - file: "max31856/figures/max31856-f020.png"
    caption: "N-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 9
  - file: "max31856/figures/max31856-f021.png"
    caption: "S-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 9
  - file: "max31856/figures/max31856-f022.png"
    caption: "K-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 9
  - file: "max31856/figures/max31856-f023.png"
    caption: "R-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 9
  - file: "max31856/figures/max31856-f024.png"
    caption: "T-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    type: "characteristic_curve"
    page: 9
  - file: "max31856/figures/max31856-f025.png"
    caption: ""
    type: "pinout"
    page: 10
  - file: "max31856/figures/max31856-f026.png"
    caption: ""
    type: "block_diagram"
    page: 11
  - file: "max31856/figures/max31856-f027.png"
    caption: "Figure 3. Serial Clock as a Function of Microcontroller Clock Polarity (CPOL)"
    type: "timing_diagram"
    page: 16
  - file: "max31856/figures/max31856-f028.png"
    caption: "Figure 4. SPI Single-Byte Read"
    type: "timing_diagram"
    page: 16
  - file: "max31856/figures/max31856-f029.png"
    caption: "Figure 5. SPI Single-Byte Write"
    type: "timing_diagram"
    page: 16
  - file: "max31856/figures/max31856-f030.png"
    caption: "Figure 6. SPI Multibyte Transfer"
    type: "timing_diagram"
    page: 17
  - file: "max31856/figures/max31856-f031.png"
    caption: "Figure 7. D̅R̅D̅Y̅ Operation"
    type: "timing_diagram"
    page: 17
  - file: "max31856/figures/max31856-f032.png"
    caption: "Figure 8. Typical Connection to Reduce the Effect of Noise Pickup in the Thermocouple Cable"
    type: "application_circuit"
    page: 27
  - file: "max31856/figures/max31856-f033.png"
    caption: "Figure 9. When Thermocouple Inputs May Be Exposed to Fault Voltages Greater than ±45V, Resistors Can be Added to Limit Current into the MAX31856."
    type: "application_circuit"
    page: 28
  - file: "max31856/figures/max31856-eq01.png"
    caption: "Equation as printed (p. 28)"
    type: equation
    page: 28
curves:
  - file: "max31856/curves/active-supply-current-vs-temperature.csv"
    caption: "ACTIVE SUPPLY CURRENT vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f006.png"
    page: 7
    x_unit: "°C"
    y_unit: "mA"
  - file: "max31856/curves/av-32-full-scale-error-vs-temperature.csv"
    caption: "AV = 32 FULL-SCALE ERROR vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f012.png"
    page: 7
    x_unit: "°C"
    y_unit: "%"
  - file: "max31856/curves/av-32-offset-error-vs-temperature.csv"
    caption: "AV = 32 OFFSET ERROR vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f016.png"
    page: 8
    x_unit: "°C"
    y_unit: "µV"
    conditions: "V_{DD} = 3.0V, 3.3V, 3.6V"
  - file: "max31856/curves/av-8-full-scale-error-vs-temperature.csv"
    caption: "AV = 8 FULL-SCALE ERROR vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f009.png"
    page: 7
    x_unit: "°C"
    y_unit: "%"
  - file: "max31856/curves/av-8-offset-error-vs-temperature.csv"
    caption: "AV = 8 OFFSET ERROR vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f013.png"
    page: 8
    x_unit: "°C"
    y_unit: "µV"
  - file: "max31856/curves/b-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "B-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f015.png"
    page: 8
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = 0°C, 25°C, 85°C, 125°C"
  - file: "max31856/curves/cold-junction-temperature-error-vs-temperature.csv"
    caption: "COLD-JUNCTION TEMPERATURE ERROR vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f017.png"
    page: 8
    x_unit: "°C"
    y_unit: "°C"
  - file: "max31856/curves/e-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "E-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f018.png"
    page: 8
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = -40°C, 25°C, 85°C,125°C"
  - file: "max31856/curves/full-scale-error-vs-common-mode-voltage.csv"
    caption: "FULL-SCALE ERROR vs. COMMON-MODE VOLTAGE"
    figure: "max31856/figures/max31856-f014.png"
    page: 8
    x_unit: "V"
    y_unit: "%"
  - file: "max31856/curves/j-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "J-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f019.png"
    page: 9
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = -40°C, 25°C, 85°C, 125°C"
  - file: "max31856/curves/k-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "K-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f022.png"
    page: 9
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = -40°C, 25°C, 85°C, 125°C"
  - file: "max31856/curves/n-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "N-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f020.png"
    page: 9
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = -40°C, 25°C, 85°C, 125°C"
  - file: "max31856/curves/r-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "R-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f023.png"
    page: 9
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = -40°C, 25°C, 85°C, 125°C"
  - file: "max31856/curves/s-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "S-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f021.png"
    page: 9
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = -40°C, 25°C, 85°C, 125°C"
  - file: "max31856/curves/standby-supply-current-vs-temperature.csv"
    caption: "STANDBY SUPPLY CURRENT vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f007.png"
    page: 7
    x_unit: "°C"
    y_unit: "µA"
  - file: "max31856/curves/t-type-linearization-error-vs-thermocouple-temperature.csv"
    caption: "T-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE"
    figure: "max31856/figures/max31856-f024.png"
    page: 9
    x_unit: "°C"
    y_unit: "°C"
    conditions: "CJ Temp = -40°C, 25°C, 85°C,125°C"
  - file: "max31856/curves/thermocouple-input-bias-current-vs-temperature.csv"
    caption: "THERMOCOUPLE INPUT BIAS CURRENT vs. TEMPERATURE"
    figure: "max31856/figures/max31856-f010.png"
    page: 7
    x_unit: "°C"
    y_unit: "nA"
    conditions: "V_{DD} = 3.0V, 3.3V, 3.6V"
  - file: "max31856/curves/thermocouple-input-differential-bias-current-vs-temperature-toc04.csv"
    caption: "THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc04)"
    figure: "max31856/figures/max31856-f008.png"
    page: 7
    x_unit: "°C"
    y_unit: "nA"
    conditions: "Differential Input Voltage = 0V, V_{DD} = 3.0V, 3.3V, 3.6V"
  - file: "max31856/curves/thermocouple-input-differential-bias-current-vs-temperature-toc05.csv"
    caption: "THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc05)"
    figure: "max31856/figures/max31856-f011.png"
    page: 7
    x_unit: "°C"
    y_unit: "nA"
    conditions: "Differential Input Voltage = 0.1V; V_{DD} = 3.0V, 3.3V, 3.6V"
agentforge:
  part_number: "MAX31856"
  datasheet_revision: "0"
  reading_order: "structure-tree"
---

# MAX31856 — Precision Thermocouple to Digital Converter with Linearization

*Datasheet: [MAX31856 data sheet (PDF)](https://www.analog.com/media/en/technical-documentation/data-sheets/max31856.pdf)*

### General Description

The  MAX31856  performs  cold-junction  compensation and digitizes  the  signal  from  any  type  of  thermocouple. The  output  data  is  formatted  in  degrees  Celsius.  This converter resolves temperatures to 0.0078125°C, allows readings  as  high  as  +1800°C  and  as  low  as  -210°C (depending on thermocouple type), and exhibits thermocouple  voltage  measurement  accuracy  of  ±0.15%.  The thermocouple  inputs  are  protected  against  overvoltage conditions up to ±45V.

A  lookup  table  (LUT)  stores  linearity  correction  data  for several types of thermocouples (K, J, N, R, S, T, E, and B). Line frequency filtering of 50Hz and 60Hz is included, as is thermocouple fault detection. A SPI-compatible interface allows selection of thermocouple type and setup of the conversion and fault detection processes.

### Applications

- Temperature Controllers
- Industrial Ovens, Furnaces, and Environmental Chambers
- Industrial Equipment

### Typical Application Circuit

![application circuit (p. 1)](max31856/figures/max31856-f000.png)

### Benefits and Features

- Provides High-Accuracy Thermocouple Temperature Readings
- Includes Automatic Linearization Correction for 8 Thermocouple Types
- ±0.15% (max, -20 ° C to +85 ° C) Thermocouple FullScale and Linearity Error
- 19-Bit, 0.0078125°C Thermocouple Temperature Resolution
- Internal Cold-Junction Compensation Minimizes System Components
- ±0.7°C (max, -20 ° C to +85 ° C) Cold-Junction Accuracy
- ±45V Input Protection Provides Robust System Performance
- Simplifies System Fault Management and Troubleshooting
-  Detects Open Thermocouples
- Over- and Undertemperature Fault Detection
- 50Hz/60Hz Noise Rejection Filtering Improves System Performance
- 14-Pin TSSOP Package

Ordering Information appears at end of data sheet.

### Absolute Maximum Ratings

**Absolute Maximum Ratings** ([table-00-p2-abs_max.csv](max31856/tables/table-00-p2-abs_max.csv), p. 2)

| PARAMETER | RATING |
|---|---|
| AVDD, DVDD | -0.3V to +4.0V |
| T+, T-, Bias | ±45V |
| T+, T-, Bias | ±20mA |
| All Other Pins | -0.3V to (V_{DVDD} + 0.3V) |
| Continuous Power Dissipation (T_{A} = +70°C) TSSOP (derate 9.1mW/°C above +70°C) | 727.3mW |
| ESD Protection (All pins, Human Body Model) | 2000V |
| Operating Temperature Range | -55°C to +125°C |
| Junction Temperature | +150°C |
| Storage Temperature Range | -65°C to +150°C |
| Lead Temperature (soldering, 10s) | +300°C |
| Soldering Temperature (reflow) | See IPC/JEDEC J-STD-020A Specification |

Stresses beyond those listed under 'Absolute Maximum Ratings' may cause permanent damage to the device. These are stress ratings only, and functional operation of the device at these or any other conditions beyond those indicated in the operational sections of the specifications is not implied. Exposure to absolute maximum rating conditions for extended periods may affect device reliability.

### (Note 1) Package Thermal Characteristics

### TSSOP

Junction-to-Ambient Thermal Resistance (θ_{JA}) — 110°C/W Junction-to-Case Thermal Resistance (θ_{JC}). — 30°C/W

- **Note 1:** Package thermal resistances were obtained using the method described in JEDEC specification JESD51-7, using a four-layer board. For detailed information on package thermal considerations, refer to www.maximintegrated.com/thermal-tutorial .

- **Recommended DC Operating Conditions** — [table-02-p2-spec.csv](max31856/tables/table-02-p2-spec.csv), p. 2

### Recommended DC Operating Conditions

(T_{A} = -55°C to +125°C, unless otherwise noted.)(Notes 2 and 4)
| PARAMETER | SYMBOL | CONDITIONS | MIN | TYP | MAX | UNITS |
|---|---|---|---|---|---|---|
| Power-Supply Voltage | V_{AVDD}, V_{DVDD} |  | ="3.0" | 3.3 | 3.6 | V |
| AVDD-DVDD |  |  | -100 |  | ="+100" | mV |
| Cable Resistance | R_{CABLE} | Per lead |  |  | 40 | kΩ |
| Input Logic 0 | V_{IL} |  |  |  | 0.8 | V |
| Input Logic 1 | V_{IH} |  | 2.1 |  |  | V |

**Electrical Characteristics** ([table-03-p2-spec.csv](max31856/tables/table-03-p2-spec.csv), p. 2)

| PARAMETER | SYMBOL | CONDITIONS | MIN | TYP | MAX | UNITS |
|---|---|---|---|---|---|---|
| Supply Current | I_{DD} | Standby |  | 5.25 | 10 | µA |
| Supply Current | I_{DD} | Active conversion |  | 1.2 | 2 | mA |
| Thermocouple Temperature Resolution |  |  |  | 19 |  | Bits |
| Thermocouple Temperature Resolution |  |  |  | 0.0078125 |  | °C |
| Cold-Junction Temperature Data Resolution |  |  |  | 0.015625 |  | °C |
| Thermocouple Input Bias Current | I_{TCBIAS} | T_{A} = +25°C | -10 |  | +10 | nA |
| Thermocouple Input Bias Current | I_{TCBIAS} | T_{A} = -40°C to 85°C | -10 |  | +65 | nA |
| Thermocouple Input Bias Current | I_{TCBIAS} | T_{A} = -55°C to +105°C | -20 |  | +110 | nA |
| Thermocouple Input Bias Current | I_{TCBIAS} | T_{A} = -55°C to +125°C | -20 |  | +400 | nA |
| Thermocouple Input Differential Bias Current (Note 4) | I_{TCIDBIAS} | T_{A} = +25°C |  | ±0.2 |  | nA |
| Thermocouple Input Differential Bias Current (Note 4) | I_{TCIDBIAS} | T_{A} = -40°C to +85°C | -4 |  | +4 | nA |
| Thermocouple Input Differential Bias Current (Note 4) | I_{TCIDBIAS} | T_{A} = -55°C to +105°C | -5.5 |  | +5.5 | nA |
| Thermocouple Input Differential Bias Current (Note 4) | I_{TCIDBIAS} | T_{A} = -55°C to +125°C | -10 |  | +10 | nA |
| Input-Referred Noise | V_{N} | AV = 8 |  | 1.3 |  | µV_{RMS} |
| Input-Referred Noise | V_{N} | AV = 32 |  | 0.4 |  | µV_{RMS} |
| Power-Supply Rejection | PSR | Cold-junction sensor |  | 0.15 |  | °C/V |
| Power-On-Reset Voltage Threshold | V_{POR} |  |  | 2.7 | 2.85 | V |
| Power-On-Reset Voltage Hysteresis | V_{HYST} |  |  | 0.1 |  | V |
| Bias Voltage | V_{BIAS} |  |  | 0.735 |  | V |
| BIAS Output Resistance | R_{BIAS} |  |  | 2 |  | kΩ |
| Input Common-Mode Range |  |  | 0.5 |  | 1.4 | V |
| Full-Scale and INL Error (Note 6) |  | T_{A} = +25°C | -0.05 |  | +0.05 | %FS |
| Full-Scale and INL Error (Note 6) |  | T_{A} = -20°C to +85°C | -0.15 |  | +0.15 | %FS |
| Full-Scale and INL Error (Note 6) |  | T_{A} = -40°C to +105°C | -0.2 |  | +0.2 | %FS |
| Full-Scale and INL Error (Note 6) |  | T_{A} = -40°C to +125°C | -0.3 |  | +0.3 | %FS |
| Full-Scale and INL Error (Note 6) |  | T_{A} = -55°C to +125°C | -0.35 |  | +0.35 | %FS |
| Input Offset Voltage (Note 7) |  | T_{A} = +25°C | -0.01 |  | +0.01 | %FS |
| Input Offset Voltage (Note 7) |  | T_{A} = -20°C to +85°C | -0.015 |  | +0.015 | %FS |
| Input Offset Voltage (Note 7) |  | T_{A} = -40°C to +105°C | -0.017 |  | +0.017 | %FS |
| Input Offset Voltage (Note 7) |  | T_{A} = -55°C to +125°C | -0.02 |  | +0.02 | %FS |
| Input Offset Voltage |  | AV = 8, T_{A} = +25°C | -7.8 |  | +7.8 | µV |
| Input Offset Voltage |  | AV = 8, T_{A} = -20°C to +85°C | -11.7 |  | +11.7 | µV |
| Input Offset Voltage |  | AV = 8, T_{A} = -40°C to +105°C | -13.3 |  | +13.3 | µV |
| Input Offset Voltage |  | AV = 8, T_{A} = -55°C to +125°C | -15.6 |  | +15.6 | µV |
| Input Offset Voltage |  | AV = 32, T_{A} = +25°C | -2.0 |  | +2.0 | µV |
| Input Offset Voltage |  | AV = 32, T_{A} = -20°C to +85°C | -2.9 |  | +2.9 | µV |
| Input Offset Voltage |  | AV = 32, T_{A} = -40°C to +105°C | -3.3 |  | +3.3 | µV |
| Input Offset Voltage |  | AV = 32, T_{A} = -55°C to +125°C | -3.9 |  | +3.9 | µV |
| Cold-Junction Temperature Error |  | T_{A} = -20°C to +85°C | -0.7 |  | +0.7 | °C |
| Cold-Junction Temperature Error |  | T_{A} = -40°C to +105°C | -1 |  | +1 | °C |
| Cold-Junction Temperature Error |  | T_{A} = -55°C to +125°C | -2 |  | +2 | °C |
| Overvoltage Rising Threshold (Note 8) |  |  | V_{AVDD} − 0.1 | V_{AVDD} + 0.17 | V_{AVDD} + 0.35 | V |
| Overvoltage Hysteresis |  |  |  | 0.09 |  | V |
| Undervoltage Falling Edge Threshold (Note 8) |  |  | -0.3 | -0.17 | 0 | V |
| Undervoltage Hysteresis |  |  |  | 0.09 |  | V |
| Thermocouple Linearity Correction Error |  | Type B, T_{A} = 0 to 125°C, T_{TC} = 95°C to +1798°C | -0.24 |  | +0.25 | °C |
| Thermocouple Linearity Correction Error |  | Type E, T_{A} = -55°C to +125°C, T_{TC} = -200°C to +1000°C | -0.14 |  | +0.06 | °C |
| Thermocouple Linearity Correction Error |  | Type J, T_{A} = -55°C to +125°C, T_{TC} = -210°C to +1200°C | -0.11 |  | +0.10 | °C |
| Thermocouple Linearity Correction Error |  | Type K, T_{A} = -55°C to +125°C, T_{TC} = -200°C to +1372°C | -0.13 |  | +0.12 | °C |
| Thermocouple Linearity Correction Error |  | Type N, T_{A} = -55°C to +125°C, T_{TC} = -200°C to +1300°C | -0.09 |  | +0.08 | °C |
| Thermocouple Linearity Correction Error |  | Type R, T_{A} = -50°C to +125°C, T_{TC} = -50°C to +1768°C | -0.19 |  | +0.17 | °C |
| Thermocouple Linearity Correction Error |  | Type S, T_{A} = -50°C to +125°C, T_{TC} = -50°C to +1768°C | -0.16 |  | +0.20 | °C |
| Thermocouple Linearity Correction Error |  | Type T, T_{A} = -55°C to +125°C, T_{TC} = -200°C to +400°C | -0.07 |  | +0.07 | °C |
| Temperature Conversion Time (Thermocouple + Cold Junction) | t_{CONV} | 1-Shot conversion or first conversion in auto-conversion mode (60Hz) |  | 143 | 155 | ms |
| Temperature Conversion Time (Thermocouple + Cold Junction) | t_{CONV} | 1-Shot conversion or first conversion in auto-conversion mode (50Hz) |  | 169 | 185 | ms |
| Temperature Conversion Time (Thermocouple + Cold Junction) | t_{CONV} | Auto conversion mode, conversions 2 through n (60Hz) |  | 82 | 90 | ms |
| Temperature Conversion Time (Thermocouple + Cold Junction) | t_{CONV} | Auto conversion mode, conversions 2 through n (50Hz) |  | 98 | 110 | ms |
| Common-Mode Rejection | CMR | 0.5V ≤ V_{CM} ≤ 1.4V |  | 70 |  | dB |
| 50/60Hz Noise Rejection |  | Fundamental and harmonics |  | 91 |  | dB |
| **SERIAL INTERFACE** |  |  |  |  |  |  |
| Input Leakage Current | I_{LEAK} | (Note 5) | -1 |  | +1 | µA |
| Output High Voltage | V_{OH} | I_{OUT} = -1.6mA | V_{CC} - 0.4 |  |  | V |
| Output Low Voltage | V_{OL} | I_{OUT} = 1.6mA |  |  | 0.4 | V |
| Input Capacitance | C_{IN} |  |  |  | 8 | pF |
| Serial Clock Frequency | f_{SCL} |  |  |  | 5 | MHz |
| SCK Pulse High Width | t_{CH} |  | 100 |  |  | ns |
| SCK Pulse Low Width | t_{CL} |  | 100 |  |  | ns |
| SCK Rise and Fall Time | t_{R}, t_{F} | C_{L} = 10pF |  |  | 200 | ns |
| C̅S̅ Fall to SCK Rise | t_{CC} | C_{L} = 10pF | 100 |  |  | ns |
| SCK to C̅S̅ Hold | t_{CCH} | C_{L} = 10pF | 100 |  |  | ns |
| C̅S̅ Rise to Output Disable | t_{CDZ} | C_{L} = 10pF |  |  | 40 | ns |
| Data to SCLK Setup | t_{DC} |  | 35 |  |  | ns |
| SCLK to Data Hold | t_{CDH} |  | 35 |  |  | ns |
| SCK Fall to Output Data Valid | t_{CDD} | C_{L} = 10pF |  |  | 80 | ns |
| C̅S̅ Inactive Time | t_{CWH} | (Note 3) | 400 |  |  | ns |

- **Note 2:** All voltages are referenced to GND. Currents entering the IC are specified positive, and currents exiting the IC are negative.

- **Note 3:** All Serial Interface timing specifications are guaranteed by design.

- **Note 4:** Specification is 100% tested at T_{A} = +25°C. Specification limits over temperature (T_{A} = T_{MIN} to T_{MAX}) are guaranteed by design and characterization; not production tested.

- **Note 5:** For all pins except T+ and T- (see the Thermocouple Input Bias Current parameter in the Electrical Characteristics table.

- **Note 6:** Using a common-mode voltage other than V_{BIAS} will change this specification. See the Typical Operating Characteristics for details.

- **Note 7:** Input-referred full-scale voltage is 78.125mV when AV = 8 and is 19.531mV when AV = 32.

- **Note 8:** Overvoltage and undervoltage limits apply to T+, T-, and BIAS pins.

![Figure 1. Timing Diagram: SPI Read Data Transfer](max31856/figures/max31856-f004.png)

*Figure 1. Timing Diagram: SPI Read Data Transfer*

![Figure 2. Timing Diagram: SPI Write Data Transfer](max31856/figures/max31856-f005.png)

*Figure 2. Timing Diagram: SPI Write Data Transfer*

![ACTIVE SUPPLY CURRENT vs. TEMPERATURE](max31856/figures/max31856-f006.png)

*ACTIVE SUPPLY CURRENT vs. TEMPERATURE*

Digitized data: [active-supply-current-vs-temperature.csv](max31856/curves/active-supply-current-vs-temperature.csv) (x: °C; y: mA) — values read off the printed plot by the digitizer, not published data.

### Typical Operating Characteristics

(V_{CC} = 3.3V and T_{A} = +25°C, unless otherwise noted.)

![STANDBY SUPPLY CURRENT vs. TEMPERATURE](max31856/figures/max31856-f007.png)

*STANDBY SUPPLY CURRENT vs. TEMPERATURE*

Digitized data: [standby-supply-current-vs-temperature.csv](max31856/curves/standby-supply-current-vs-temperature.csv) (x: °C; y: µA) — values read off the printed plot by the digitizer, not published data.

![THERMOCOUPLE INPUT BIAS CURRENT vs. TEMPERATURE](max31856/figures/max31856-f010.png)

*THERMOCOUPLE INPUT BIAS CURRENT vs. TEMPERATURE*

Digitized data: [thermocouple-input-bias-current-vs-temperature.csv](max31856/curves/thermocouple-input-bias-current-vs-temperature.csv) (x: °C; y: nA) — values read off the printed plot by the digitizer, not published data.

![THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc04)](max31856/figures/max31856-f008.png)

*THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc04)*

Digitized data: [thermocouple-input-differential-bias-current-vs-temperature-toc04.csv](max31856/curves/thermocouple-input-differential-bias-current-vs-temperature-toc04.csv) (x: °C; y: nA) — values read off the printed plot by the digitizer, not published data.

![THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc05)](max31856/figures/max31856-f011.png)

*THERMOCOUPLE INPUT DIFFERENTIAL BIAS CURRENT vs. TEMPERATURE (toc05)*

Digitized data: [thermocouple-input-differential-bias-current-vs-temperature-toc05.csv](max31856/curves/thermocouple-input-differential-bias-current-vs-temperature-toc05.csv) (x: °C; y: nA) — values read off the printed plot by the digitizer, not published data.

![AV = 8 FULL-SCALE ERROR vs. TEMPERATURE](max31856/figures/max31856-f009.png)

*AV = 8 FULL-SCALE ERROR vs. TEMPERATURE*

Digitized data: [av-8-full-scale-error-vs-temperature.csv](max31856/curves/av-8-full-scale-error-vs-temperature.csv) (x: °C; y: %) — values read off the printed plot by the digitizer, not published data.

![AV = 32 FULL-SCALE ERROR vs. TEMPERATURE](max31856/figures/max31856-f012.png)

*AV = 32 FULL-SCALE ERROR vs. TEMPERATURE*

Digitized data: [av-32-full-scale-error-vs-temperature.csv](max31856/curves/av-32-full-scale-error-vs-temperature.csv) (x: °C; y: %) — values read off the printed plot by the digitizer, not published data.

![AV = 8 OFFSET ERROR vs. TEMPERATURE](max31856/figures/max31856-f013.png)

*AV = 8 OFFSET ERROR vs. TEMPERATURE*

Digitized data: [av-8-offset-error-vs-temperature.csv](max31856/curves/av-8-offset-error-vs-temperature.csv) (x: °C; y: µV) — values read off the printed plot by the digitizer, not published data.

![AV = 32 OFFSET ERROR vs. TEMPERATURE](max31856/figures/max31856-f016.png)

*AV = 32 OFFSET ERROR vs. TEMPERATURE*

Digitized data: [av-32-offset-error-vs-temperature.csv](max31856/curves/av-32-offset-error-vs-temperature.csv) (x: °C; y: µV) — values read off the printed plot by the digitizer, not published data.

### Typical Operating Characteristics (continued)

(V_{CC} = 3.3V and T_{A} = +25°C, unless otherwise noted.)

![FULL-SCALE ERROR vs. COMMON-MODE VOLTAGE](max31856/figures/max31856-f014.png)

*FULL-SCALE ERROR vs. COMMON-MODE VOLTAGE*

Digitized data: [full-scale-error-vs-common-mode-voltage.csv](max31856/curves/full-scale-error-vs-common-mode-voltage.csv) (x: V; y: %) — values read off the printed plot by the digitizer, not published data.

![COLD-JUNCTION TEMPERATURE ERROR vs. TEMPERATURE](max31856/figures/max31856-f017.png)

*COLD-JUNCTION TEMPERATURE ERROR vs. TEMPERATURE*

Digitized data: [cold-junction-temperature-error-vs-temperature.csv](max31856/curves/cold-junction-temperature-error-vs-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![B-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f015.png)

*B-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [b-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/b-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![E-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f018.png)

*E-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [e-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/e-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![J-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f019.png)

*J-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [j-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/j-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![K-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f022.png)

*K-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [k-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/k-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![N-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f020.png)

*N-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [n-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/n-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![R-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f023.png)

*R-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [r-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/r-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![S-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f021.png)

*S-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [s-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/s-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

![T-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE](max31856/figures/max31856-f024.png)

*T-TYPE LINEARIZATION ERROR vs. THERMOCOUPLE TEMPERATURE*

Digitized data: [t-type-linearization-error-vs-thermocouple-temperature.csv](max31856/curves/t-type-linearization-error-vs-thermocouple-temperature.csv) (x: °C; y: °C) — values read off the printed plot by the digitizer, not published data.

### Pin Configuration

![pinout (p. 10)](max31856/figures/max31856-f025.png)

### Pin Description

**Pin Description** ([table-04-p10-pin.csv](max31856/tables/table-04-p10-pin.csv), p. 10)

| PIN | NAME | FUNCTION |
|---|---|---|
| 1 | AGND | Analog Ground |
| 2 | BIAS | Bias Voltage Source. Nominally 0.735V. This pin is floating when no conversions are taking place. |
| 3 | T- | Thermocouple Negative Input. See Table 1. |
| 4 | T+ | Thermocouple Positive Input. See Table 1. |
| 5 | AVDD | Analog Positive Supply. Bypass with a 0.1µF capacitor to AGND. |
| 6 | DNC | Do Not Connect |
| 7 | D̅R̅D̅Y̅ | Data Ready Output |
| 8 | DVDD | Digital Positive Supply. Bypass with a 0.1µF capacitor to DGND. |
| 9 | C̅S̅ | Chip Select. Set C̅S̅ low to enable the serial interface. |
| 10 | SCK | Serial Clock Input |
| 11 | SDO | Serial Data Output |
| 12 | SDI | Serial Data Input |
| 13 | F̅A̅U̅L̅T̅ | Cable, thermocouple, or temperature fault output |
| 14 | DGND | Digital Ground |

### Block Diagram

![block diagram (p. 11)](max31856/figures/max31856-f026.png)

### Detailed Description

The MAX31856 is a sophisticated thermocouple-to-digital converter with a built-in 19-bit analog-to-digital converter (ADC).  Internal  functions  include  correction  for  thermocouple  nonlinearity,  input  protection,  cold-junction  compensation  sensing  and  correction,  a  digital  controller,  a SPI-compatible interface, and associated control logic.

In the simplest configuration, the thermocouple wires connect  directly  to  inputs  T-  and  T+,  with  a  common-mode bias  voltage  provided  by  the  BIAS  output.  Additional filtering  and/or  protection  components  may  be  added  if needed,  as  discussed  in  the Applications  Information section. Operation is controlled by two configuration bytes and four bytes that contain over- and undertemperature detection thresholds.

### Temperature Conversion

The  temperature  conversion  process  consists  of  five steps as described in the sections below. The input amplifier and ADC amplify and digitize the thermocouple's voltage  output.  The  internal  temperature  sensor  measures the  cold-junction  temperature.  Using  the  internal  lookup table  (LUT),  the  ADC  code  corresponding  to  the  coldjunction temperature for the selected thermocouple type is determined. The thermocouple code and the cold-junction code are summed to produce the code corresponding to the cold-junction compensated thermocouple temperature. Finally, the LUT is used to produce a cold-junction compensated output code in units of °C.

### Thermocouple Voltage Conversion

T+  and  T-  are  the  thermocouple  inputs.  T-  is  biased  to approximately 0.735V by the BIAS output. The amplifier provides gain to the µV- and mV-level thermocouple sig -nals to make the amplitude appropriate for the ADC's fullscale input range. Two amplifier gains provide full-scale input ranges of ±78.125mV and ±19.531mV to accommodate higher- and lower-sensitivity thermocouples.

Because  long  thermocouple  wires  may  pick  up  noise from a variety of sources, including AC power cables, the amplified  signal  is  lowpass  filtered  before  being  applied to the ADC. The ADC provides further digital lowpass and notch filtering to attenuate input noise. The notch frequencies  are  either  50Hz  and  its  harmonics  or  60Hz  and  its harmonics, selectable using bit 0 of the Configuration 0 register (00h). In addition, bits D6:4 of the Configuration 1 register (01h) enable an averaging mode that provides additional filtering with an associated increase in conversion time. 2, 4, 8, or 16 samples may be averaged using this mode.

The conversion mode can be either continuous or 'normally  off',  as  selected  by  bit  7  of  the  Configuration  0 register  (00h).  When  in  the  normally  off  mode,  a  single '1-shot'  conversion  may  be  selected  using  bit  6  of  the Configuration 0 register (00h).

Thermocouple type is user-selectable using bits D3:0 of the  Configuration  1  register  (01h).  Thermocouple  types K,  J,  N,  R,  S,  T,  B,  and  E  are  supported  by  automatic cold-junction  compensation  and  linearization.  (To  use a different  thermocouple type, use bits D3:0 to select a gain of either 8 or 32. The linearization and cold-junction compensation calculations may then be done externally using  the  cold-junction  temperature  and  thermocouple voltage data.)

### Cold-Junction Temperature Sensing

The function of the thermocouple is to sense a difference in temperature between two ends of the thermocouple wires. The thermocouple's sensing junction (often called the 'hot' junction  regardless  of  its  temperature)  can  be  measured across its rated operating temperature range (see Table 1 for supported thermocouple temperature ranges).

Additional thermocouples are created where the thermocouple wires make contact with different metals, usually at a connector or at the point where they are soldered to a PCB (the 'cold junction'). To compensate for the errors due to these additional thermocouples, the temperature at the cold junction must be measured. This is done with the internal precision temperature sensor, which has accuracy better  than  ±0.7°C  from  -20°C  to  +85°C.  By  placing  the MAX31856 near the cold junction, the cold-junction temperature  can  be  measured  and  used  to  compensate  for cold-junction effects.

The MAX31856 stores the cold-junction temperature data in  registers  0Ah  and  0Bh.  When  the  cold-junction  temperature sensor is enabled, these registers are read-only and contain the measured cold-junction temperature plus the  value  in  the  Cold-Junction  Offset  register.  Reading the  register  with  the  cold-junction  temperature  sensor enabled will reset the D̅R̅D̅Y̅ pin high. Both bytes of this register should be read as a multibyte transfer to ensure both bytes are from the same temperature update. When the  cold-junction  temperature  sensor  is  disabled,  these registers  become  read-write  registers  that  contain  the most recent measured temperature value. If desired, data from  an  external  temperature  sensor  may  be  written  to these  registers  when  the  internal  cold-junction  sensor is  disabled.  The  maximum  cold-junction  temperature  is clamped at 128°C and the minimum is clamped at -64°C. See Table 2 for the Reference Junction (Cold Junction) Temperature Data Format.

If  desired,  a  temperature  offset  may  be  written  to  the Cold-Junction  Offset  register  (09h).  The  value  stored  in registers 0Ah and 0Bh will then be equal to the measured

### Table 1. Supported Thermocouples and Temperature Ranges

**Table 1. Supported Thermocouples and Temperature Ranges** ([table-05-p12-data.csv](max31856/tables/table-05-p12-data.csv), p. 12)

| TYPE | T-WIRE | T+ WIRE | TEMP RANGE | NOMINAL SENSITIVITY (µV/°C) | COLD-JUNCTION TEMP RANGE |
|---|---|---|---|---|---|
| B | Platinum/Rhodium | Platinum/Rhodium | 250^{o}C to 1820^{o}C | 10.086 (+500°C to +1500°C) | 0 to 125°C |
| E | Constantan | Chromel | -200°C to +1000°C | 76.373 (0°C to +1000°C) | -55°C to +125°C |
| J | Constantan | Iron | -210°C to +1200°C | 57.953 (0°C to + 750°C) | -55°C to +125°C |
| K | Alumel | Chromel | -200°C to +1372°C | 41.276 (0°C to + 1000°C) | -55°C to +125°C |
| N | Nisil | Nicrosil | -200°C to +1300°C | 36.256 (0°C to +1000°C) | -55°C to +125°C |
| R | Platinum | Platinum/Rhodium | -50°C to +1768°C | 10.506 (0°C to +1000°C) | -50°C to +125°C |
| S | Platinum | Platinum/Rhodium | -50°C to +1768°C | 9.587 (0°C to +1000°C) | -50°C to +125°C |
| T | Constantan | Copper | -200°C to +400°C | 52.18 (0°C to +400°C) | -55°C to +125°C |

value plus the offset value. The MSB of the offset register is  4°C and the LSB is 0.0625°C. The resulting range of the offset value applied to the measured CJ temperature is  -8°C  to  +7.9375°C.  The  default  offset  value  is  0°C (00h).

### Thermocouple Linearization and Conversion of Code to Temperature

Optimal performance is achieved when the thermocouple cold junction and the cold-junction sensor are at the same temperature.  Avoid  placing  heat-generating  devices  or components near the cold junction because this may produce cold-junction-related errors. When a significant temperature  differential  between  the  internal  sensor  and  the cold junction is unavoidable, an external temperature sensor may be used instead. The temperature measured by the external sensor may be written to the cold-junction temperature register and used for cold-junction compensation. Bit 3 of Configuration 0 register (00h) disables the internal cold-junction temperature sensor and allows temperature values from an external sensor to be written directly into the Cold-Junction Temperature registers (0Ah and 0Bh).

### Cold-Junction Temperature Translation and Compensation

Thermocouple  temperature  values  and  corresponding ADC codes are stored in an internal lookup table. After measuring  the  cold-junction  temperature,  the  LUT  is used to convert the temperature value to the equivalent ADC  code  for  the  type  of  thermocouple  being  used. Values between LUT entries are interpolated. The coldjunction ADC  code  is  added  to  the  conversion  result  in the thermocouple voltage register to yield a cold-junctioncompensated value.

### Table 2. Reference Junction (Cold-Junction) Temperature Data Format

**Table 2. Reference Junction (Cold-Junction) Temperature Data Format** ([table-06-p13-data.csv](max31856/tables/table-06-p13-data.csv), p. 13)

| TEMPERATURE (°C) | DIGITAL OUTPUT |
|---|---|
| +127.984375 | 0111 1111 1111 1100 |
| +127 | 0111 1111 0000 0000 |
| +125 | 0111 1101 0000 0000 |
| +64 | 0100 0000 0000 0000 |
| +25 | 0001 1001 0000 0000 |
| +0.5 | 0000 0000 1000 0000 |
| +0.015625 | 0000 0000 0000 0100 |
| 0 | 0000 0000 0000 0000 |
| -0.5 | 1111 1111 1000 0000 |
| -25 | 1110 0111 0000 0000 |
| -55 | 1100 1001 0000 0000 |

### Thermocouple Linearization and Conversion of Code to Temperature 

Because all thermocouples are nonlinear, the raw coldjunction-compensated value must be corrected for nonlinearity  and  converted  to  a  temperature  value.  This is  done  using  the  LUT  to  produce  the  linearized  and cold-junction-compensated  temperature  value,  which  is stored after every conversion as 19 bits in the Linearized Thermocouple  Temperature  registers  (0Ch,  0Dh,  and 0Eh). All three bytes should be read as a multibyte transfer  to  ensure  all  are  from  the  same  data  update.  See Table  3  for  the  Linearized  Thermocouple  Temperature Data Format.

Linearization accuracy varies by thermocouple type, 'hotjunction' temperature, and cold-junction temperature, with the largest errors typically occurring near the hot-junction and cold-junction extremes. Worst-case values for linearization errors are shown in the Electrical Characteristics table.

### Over-/Undertemperature Fault Detection

Over- and undertemperature fault detection are available for both the cold-junction temperature and the linearized and cold-junction-compensated temperature reading. Two registers (03h and 04h) contain the high and low thresholds for the cold-junction temperature. The cold-junction temperature value in registers 0Ah and 0Bh is compared to  the  threshold  values.  If  a  threshold  is  exceeded,  the corresponding bit is set in the Fault Status register (0Fh) and, if not masked, the F̅A̅U̅L̅T̅ output will assert.

### Table 3. Linearized Thermocouple Temperature Data Format

**Table 3. Linearized Thermocouple Temperature Data Format** ([table-07-p13-data.csv](max31856/tables/table-07-p13-data.csv), p. 13)

| TEMPERATURE (°C) | DIGITAL OUTPUT |
|---|---|
| +1600.00 | 0110 0100 0000 0000 0000 0000 |
| +1000.00 | 0011 1110 1000 0000 0000 0000 |
| +100.9375 | 0000 0110 0100 1111 0000 0000 |
| +25.00 | 0000 0001 1001 0000 0000 0000 |
| +0.0625 | 0000 0000 0000 0001 0000 0000 |
| 0.00 | 0000 0000 0000 0000 0000 0000 |
| -0.0625 | 1111 1111 1111 1111 0000 0000 |
| -0.25 | 1111 1111 1111 1100 0000 0000 |
| -1.00 | 1111 1111 1111 0000 0000 0000 |
| -250.00 | 1111 0000 0110 0000 0000 0000 |

- This format also applies to the High Fault and Low Fault thresholds.
- (Note that the practical temperature range varies with the thermocouple type.)

Four registers (05h through 08h) contain over- and undertemperature thresholds for the linearized and cold-junction-compensated temperature. These threshold register values are compared to the linearized temperature reading found in registers 0Ch, 0Dh, and 0Eh. If a threshold is exceeded, the corresponding bit is set in the Fault Status register (0Fh) and, if not masked, the F̅A̅U̅L̅T̅ output will assert.

### Integrated Input Protection

The  internal  circuitry  is  protected  from  excessive  voltages  applied  to  the  thermocouple  cables  by  integrated MOSFETs at the T+ and T- inputs, and the BIAS output. These MOSFETs turn off when the input voltage is negative or greater than V_{DD}. The MOSFETs are capable of withstanding input voltages up to ±45V. If fault voltages beyond the ±45V limits are expected, see the Applications Information section.

When the absolute input voltage at T+ or T- is negative or greater than V_{DD}, the Under-/Overvoltage Fault bit, Bit 1, is set in the Fault Status register (0Fh) and the F̅A̅U̅L̅T̅ pin asserts if not masked. Conversions are suspended while the OVUV fault is present and will resume when the fault is removed.

### Open-Circuit Fault Detection

Detection  of  open-circuit  faults,  such  as  those  caused by  broken  thermocouple  wires,  can  be  enabled  or  disabled using bits 4 and 5 in the Configuration 0 register (00h). Fault detection is accomplished by forcing a small

### Table 4. Open-Circuit Detection Mode

**Table 4. Open-Circuit Detection Mode** ([table-08-p14-data.csv](max31856/tables/table-08-p14-data.csv), p. 14)

*Column groups as printed — columns 4–5: FAULT TEST TIME (ms) CJ SENSE ENABLED · columns 6–7: FAULT TEST TIME (ms) CJ SENSE DISABLED.*

| BITS 5:4 OCFAULT1: OCFAULT0 (Config Byte 0) | FAULT TEST | INPUT NETWORK | TYP | MAX | TYP | MAX |
|---|---|---|---|---|---|---|
| 00 | Disabled | N/A | 0 | 0 | 0 | 0 |
| 01 | Enabled (Once every 16 conversions) | R_{S} < 5kΩ | 13.3 | 15 | 40 | 44 |
| 10 | Enabled (Once every 16 conversions) | 40kΩ > R_{S} > 5kΩ; Time constant < 2ms | 33.4 | 37 | 60 | 66 |
| 11 | Enabled (Once every 16 conversions) | 40kΩ > R_{S} > 5kΩ; Time constant > 2ms | 113.4 | 125 | 140 | 154 |

current through the thermocouple wires. The time required to  detect  an  open  circuit  depends  on  the  values  of  the lead resistances and any filter capacitance at the thermocouple input and therefore, bits 4 and 5 also select the time  allowed  for  open-circuit  fault  detection.  A  nominal detection  time  of  either  10ms,  32ms,  or  100ms  can  be selected. The  Open-Circuit Detection  Mode  table  (T able 4) shows the effect of these two  bits on the conversion time. When the device is in one-shot mode, open-circuit detection can be disabled or set to occur every one-shot conversion. When the device is in automatic conversion mode, open-circuit  detection  may  be  disabled,  or  it may be set to automatically test for open circuits every 16 conversion  cycles. If on-demand detection is desired, select 'detection  disabled'  (00),  then  select  the  setting  for  the desired time constant. An open-circuit detection test will be performed immediately after the current conversion is completed. Disabling the open fault detection when in comparator mode while there is an open fault present will not clear the fault bit or F̅A̅U̅L̅T̅ pin. If this happens, to  subsequently  clear  the  fault,  the  MAX31856  must be placed in interrupt mode and then the fault cleared. Note that, when cold-junction sensing is enabled, opencircuit  fault  detection  and  cold-junction  sensing  occur concurrently. Therefore,  cold-junction  temperature  sensing  has  no  effect  on  the  overall  cycle  time  when  opencircuit fault detection is enabled.  An open-circuit fault is indicated by the Open Fault bit, Bit 0, in the Fault Status register (0Fh) and the F̅A̅U̅L̅T̅ pin asserts if not masked.

### Cold-Junction and Thermocouple Out-of-Range Detection

Thermocouple characteristics, the measurement circuitry, and the linearization calculations limit the optimum temperature ranges for both the cold junction and the measurement  junction  ('hot  junction').  Bit  D7  of  the  Fault Status register indicates when the cold-junction temperature falls outside of the optimum range, and bit D6 indicates when the hot-junction temperature is out of range. Table  1  shows  the  temperature  limits  that  apply  for  the supported thermocouple types. These values are rounded to the nearest °C. When the temperature falls outside of the limit for a given measurement, the reported thermocouple  temperature  is  clamped  at  the  limit  value.  Note that the F̅A̅U̅L̅T̅ pin never asserts for an out-of-range fault.

### Serial Interface

Four pins are used for SPI-compatible communications: SDO  (serial-data  out),  SDI  (serial-data  in), C̅S̅ (chip select),  and  SCLK  (serial  clock).  SDI  and  SDO  are  the serial-data  input  and  output  pins,  respectively.  The C̅S̅ input initiates and terminates a data transfer. SCLK synchronizes data movement between the master (microcontroller) and the slave (MAX31856).

The  serial  clock  (SCLK),  which  is  generated  by  the microcontroller,  is  active  only  when C̅S̅ is  low  and  during address and data transfer to any device on the SPI bus. The inactive clock polarity is programmable in some microcontrollers.  The  MAX31856  automatically  accommodates either clock polarity by sampling SCLK when C̅S̅ becomes active to determine the polarity of the inactive clock.  Input  data  (SDI)  is  latched  on  the  internal  strobe edge  and  output  data  (SDO)  is  shifted  out  on  the  shift edge (see Table 5 and Figure 3). There is one clock for each bit transferred. Address and data bits are transferred in groups of eight, MSB first. 

### Table 5. Serial Interface Function

**Table 5. Serial Interface Function** ([table-09-p15-data.csv](max31856/tables/table-09-p15-data.csv), p. 15)

| MODE | C̅S̅ | SCLK | SDI | SDO |
|---|---|---|---|---|
| Disable Reset | High | Input Disabled | Input disabled | High impedance |
| Write | Low | CPOL = 1*, SCLK rising | Data bit latch | High impedance |
| Write | Low | CPOL = 0, SCLK falling | Data bit latch | High impedance |
| Read | Low | CPOL = 1, SCLK falling | X | Next data bit shift** |
| Read | Low | CPOL = 0, SCLK rising | X | Next data bit shift** |

- Note: CPHA bit polarity must be set to 1.
- **\*:** CPOL is the clock polarity bit that is set in the control register of the microcontroller.
- **\*\*:** SDO remains at high impedance until 8 bits of data are ready to be shifted out during a read.

### Address and Data Bytes

Address  and  data  bytes  are  shifted  MSB-first  into  the serial-data  input  (SDI)  and  out  of  the  serial-data  output (SDO). Any transfer requires the address of the byte to specify a write or a read, followed by one or more bytes of data. Data is transferred out of the SDO for a read operation and into the SDI for a write operation. The address byte is always the first byte transferred after C̅S̅ is driven low. The MSB (A7) of this byte determines whether the following  byte  will  be  written  or  read.  If A7  is  0,  one  or more byte reads will follow the address byte. If A7 is 1, one or more byte writes will follow the address byte.

For a single-byte transfer,  1  byte  is  read  or  written  and then C̅S̅ is  driven high (see Figure 4 and Figure 5). For a  multiple-byte  transfer,  multiple  bytes  can  be  read  or written after the address has been written (see Figure 6). The address continues to increment through all memory locations as long as C̅S̅ remains low. If data continues to be clocked in or out, the address will loop from 7Fh/FFh to 00h/80h. Invalid memory addresses report an FFh value. Attempting to write to a read-only register will result in no change to that register's contents.

### D̅R̅D̅Y̅

The D̅R̅D̅Y̅ output goes low when a new conversion result is available in the Linearized Thermocouple Temperature register. When  a  read-operation of the Linearized Thermocouple Temperature register or the Cold-Junction Temperature  Register  (if  enabled)  completes, D̅R̅D̅Y̅ returns high.

![Figure 3. Serial Clock as a Function of Microcontroller Clock Polarity (CPOL)](max31856/figures/max31856-f027.png)

*Figure 3. Serial Clock as a Function of Microcontroller Clock Polarity (CPOL)*

![Figure 4. SPI Single-Byte Read](max31856/figures/max31856-f028.png)

*Figure 4. SPI Single-Byte Read*

![Figure 5. SPI Single-Byte Write](max31856/figures/max31856-f029.png)

*Figure 5. SPI Single-Byte Write*

![Figure 6. SPI Multibyte Transfer](max31856/figures/max31856-f030.png)

*Figure 6. SPI Multibyte Transfer*

![Figure 7. D̅R̅D̅Y̅ Operation](max31856/figures/max31856-f031.png)

*Figure 7. D̅R̅D̅Y̅ Operation*

### Internal Registers

### Address and Data Bytes

Communication with  the  MAX31856  is  accomplished  through  16  8-bit  registers  that  contain  conversion,  status,  and configuration data. All programming is done by selecting the appropriate address of the desired register location. The Register Memory Map (Table 6) illustrates the addresses for the temperature, status, and configuration registers.

The registers are accessed using the 0Xh addresses for reads and the 8Xh addresses for writes. Data is read from or written to the registers MSB first. Attempts to write to a read-only register results in no change in the data.

### Table 6. Register Memory Map

**Table 6. Register Memory Map** ([table-10-p18-register.csv](max31856/tables/table-10-p18-register.csv), p. 18)

| ADDRESS | READ/WRITE | NAME | FACTORY DEFAULT | FUNCTION |
|---|---|---|---|---|
| 00h/80h | Read/Write | CR0 | 00h | Configuration 0 Register |
| 01h/81h | Read/Write | CR1 | 03h | Configuration 1 Register |
| 02h/82h | Read/Write | MASK | FFh | Fault Mask Register |
| 03h/83h | Read/Write | CJHF | 7Fh | Cold-Junction High Fault Threshold |
| 04h/84h | Read/Write | CJLF | C0h | Cold-Junction Low Fault Threshold |
| 05h/85h | Read/Write | LTHFTH | 7Fh | Linearized Temperature High Fault Threshold MSB |
| 06h/86h | Read/Write | LTHFTL | FFh | Linearized Temperature High Fault Threshold LSB |
| 07h/87h | Read/Write | LTLFTH | 80h | Linearized Temperature Low Fault Threshold MSB |
| 08h/88h | Read/Write | LTLFTL | 00h | Linearized Temperature Low Fault Threshold LSB |
| 09h/89h | Read/Write | CJTO | 00h | Cold-Junction Temperature Offset Register |
| 0Ah/8Ah | Read/Write | CJTH | 00h | Cold-Junction Temperature Register, MSB |
| 0Bh/8Bh | Read/Write | CJTL | 00h | Cold-Junction Temperature Register, LSB |
| 0Ch | Read Only | LTCBH | 00h | Linearized TC Temperature, Byte 2 |
| 0Dh | Read Only | LTCBM | 00h | Linearized TC Temperature, Byte 1 |
| 0Eh | Read Only | LTCBL | 00h | Linearized TC Temperature, Byte 0 |
| 0Fh | Read Only | SR | 00h | Fault Status Register |

### Register 00h/80h: Configuration 0 Register (CR0)

The Configuration 0 register selects the conversion mode (automatic or triggered by the 1-shot command), selects opencircuit fault detection timing, enables the cold-junction sensor, clears the fault status register, and selects the filter notch frequencies. The effects of the configuration bits are described below.

Default Value: 00h

**Register 00h/80h: Configuration 0 Register (CR0) bit map** ([table-11-p18-register.csv](max31856/tables/table-11-p18-register.csv), p. 18)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 00h/80h | CMODE | 1SHOT | OCFAULT1 | OCFAULT0 | CJ | FAULT | FAULTCLR | 50/60Hz |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 00h/80h: Configuration 0 Register (CR0) (continued)

**Register 00h/80h: Configuration 0 Register (CR0) (continued) — Bit Descriptions** ([table-12-p19-register.csv](max31856/tables/table-12-p19-register.csv), p. 19)

| BIT | NAME | DESCRIPTION |
|---|---|---|
| 7 | CMODE | Conversion Mode 0 = Normally Off mode (default) 1 = Automatic Conversion mode. Conversions occur continuously every 100ms (nominal). |
| 6 | 1SHOT | One-Shot Mode 0 = No conversions requested (default) 1 = This causes a single cold-junction and thermocouple conversion to take place when Conversion Mode bit =0 (normally off mode). The conversion is triggered when C̅S̅ goes high after writing a 1 to this bit. Note that if a multi-byte write is performed, the conversion is triggered when C̅S̅ goes high at the end of the transaction. A single conversion requires approximately 143ms in 60Hz filter mode or 169ms in 50Hz filter mode to complete. This bit self clears to 0. |
| 5:4 | OCFAULT[1:0] | These bits enable/disable open-circuit fault detection and select fault detection timing. See Open-Circuit Fault Detection section and Table 4 for operation of these bits. |
| 3 | CJ | Cold-Junction Sensor Disable 0 = Cold-junction temperature sensor enabled (default) 1 = Cold-junction temperature sensor disabled. Data from an external temperature sensor may be written to the cold-junction temperature register. When this bit changes from 0 to 1, the most recent cold-junction temperature value will remain in the cold-junction temperature register until the internal sensor is enabled or until a new value is written to the register. The overall temperature conversion time is reduced by 25ms (typ) when this bit is set to 1. |
| 2 | FAULT | Fault Mode 0 = Comparator Mode. The F̅A̅U̅L̅T̅ output and respective fault bit reflects the state of any non-masked faults by asserting when the fault condition is true, and deasserting when the fault condition is no longer true. There is a 2°C hysteresis when in comparator mode for threshold fault conditions. (default) 1 = Interrupt Mode. The F̅A̅U̅L̅T̅ output and respective fault bit asserts when a non-masked fault condition is true and remain asserted until a 1 is written to the Fault Status Clear bit. This deasserts F̅A̅U̅L̅T̅ and respective fault bit until a new fault is detected (note that this may occur immediately if the fault condition is still in place). |
| 1 | FAULTCLR | Fault Status Clear 0 = Default 1 = When in interrupt mode, returns all Fault Status bits [7:0] in the Fault Status Register (0Fh) to 0 and deasserts the F̅A̅U̅L̅T̅ output. This bit has no effect in comparator mode. Note that the F̅A̅U̅L̅T̅ output and the fault bit may reassert immediately if the fault persists. To prevent the F̅A̅U̅L̅T̅ output from reasserting, first set the Fault Mask bits. The fault status clear bit self-clears to 0. |
| 0 | 50/60Hz | 50Hz/60Hz Noise Rejection Filter Selection 0= Selects rejection of 60Hz and its harmonics (default) 1= Selects rejection of 50Hz and its harmonics Note: Change the notch frequency only while in the “Normally Off” mode – not in the Automatic Conversion mode. |

### Register 01h/81h: Configuration 1 Register (CR1)

The Configuration 1 register selects the averaging time for the thermocouple voltage conversion averaging mode and also selects the thermocouple type being monitored.

Default Value: 03h

**Register 01h/81h: Configuration 1 Register (CR1) bit map** ([table-13-p20-register.csv](max31856/tables/table-13-p20-register.csv), p. 20)

| MEMORY ACCESS | N/A | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 01h/81h | Reserved | AVGSEL_{2} | AVGSEL_{1} | AVGSEL_{0} | TC TYPE_{3} | TC TYPE_{2} | TC TYPE_{1} | TC TYPE_{0} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

**Register 01h/81h: Configuration 1 Register (CR1) — Bit Descriptions** ([table-14-p20-register.csv](max31856/tables/table-14-p20-register.csv), p. 20)

| BIT | NAME | DESCRIPTION |
|---|---|---|
| 7 | Reserved | Reserved. |
| 6:4 | AVGSEL[2:0] | Thermocouple Voltage Conversion Averaging Mode 000 = 1 sample (default) 001 = 2 samples averaged 010 = 4 samples averaged 011 = 8 samples averaged 1xx = 16 samples averaged Adding samples increases the conversion time and reduces noise. Typical conversion times: 1-shot or first conversion in Auto mode: = t_{CONV} + (samples -1) x 33.33mS (60Hz rejection) = t_{CONV} + (samples -1) x 40mS (50Hz rejection) 2 thru n conversions in Auto mode = t_{CONV} + (samples -1) x 16.67mS (60Hz rejection) = t_{CONV} + (samples -1) x 20mS (50Hz rejection)  The Thermocouple Voltage Conversion Averaging Mode settings should not be changed while conversions are taking place. |
| 3:0 | TC TYPE[3:0] | Thermocouple Type 0000 = B Type 0001 = E Type 0010 = J Type 0011 = K Type (default) 0100 = N Type 0101 = R Type 0110 = S Type 0111 = T Type 10xx = Voltage Mode, Gain = 8. Code = 8 x 1.6 x 2^{17} x V_{IN} 11xx = Voltage Mode, Gain = 32. Code = 32 x 1.6 x 2^{17} x V_{IN} Where Code is 19 bit signed number from TC registers and V_{IN} is thermocouple input voltage |

### Register 02h/82h: Fault Mask Register (MASK)

The Fault Mask Register allows the user to mask faults from causing the F̅A̅U̅L̅T̅ output from asserting. Masked faults will still result in fault bits being set in the Fault Status register (0Fh). Note that the F̅A̅U̅L̅T̅ output is never asserted by thermocouple and cold-junction out-of-range status.

### Default Value: FFh

**Register 02h/82h: Fault Mask Register (MASK) bit map** ([table-15-p21-register.csv](max31856/tables/table-15-p21-register.csv), p. 21)

| MEMORY ACCESS | N/A | N/A | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 02h/82h | Reserved | Reserved | CJ High FAULT Mask | CJ Low FAULT Mask | TC High FAULT Mask | TC Low FAULT Mask | OV/UV FAULT Mask | Open FAULT Mask |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

**Register 02h/82h: Fault Mask Register (MASK) — Bit Descriptions** ([table-16-p21-register.csv](max31856/tables/table-16-p21-register.csv), p. 21)

| BIT | NAME | DESCRIPTION |
|---|---|---|
| 7:6 | Reserved | Reserved. |
| 5 | CJ High FAULT Mask | Cold-Junction High Fault Threshold Mask 0 = F̅A̅U̅L̅T̅ output asserted when the Cold-Junction Temperature rises above the Cold-Junction Temperature high threshold limit value 1 = F̅A̅U̅L̅T̅ output masked (default) |
| 4 | CJ Low FAULT Mask | Cold-Junction Low Fault Threshold Mask 0 = F̅A̅U̅L̅T̅ output asserted when the Cold-Junction Temperature falls below the Cold-Junction Temperature low threshold limit value 1 = F̅A̅U̅L̅T̅ output masked (default) |
| 3 | TC High FAULT Mask | Thermocouple Temperature High Fault Threshold Mask 0 = F̅A̅U̅L̅T̅ output asserted when the Thermocouple Temperature rises above the Thermocouple Temperature high threshold limit value 1 = F̅A̅U̅L̅T̅ output masked (default) |
| 2 | TC Low FAULT Mask | Thermocouple Temperature Low Fault Threshold Mask 0 = F̅A̅U̅L̅T̅ output asserted when the Thermocouple Temperature falls below the Thermocouple Temperature low threshold limit value 1 = F̅A̅U̅L̅T̅ output masked (default) |
| 1 | OV/UV FAULT Mask | Over-voltage or Undervoltage Input Fault Mask 0 = F̅A̅U̅L̅T̅ output asserted when an over- or undervoltage condition is detected 1 = F̅A̅U̅L̅T̅ output masked (default) |
| 0 | Open FAULT Mask | Thermocouple Open-Circuit Fault Mask 0 = F̅A̅U̅L̅T̅ output asserted when a thermocouple open condition is detected 1 = F̅A̅U̅L̅T̅ output masked (default) |

### Register 03h/83h: Cold-Junction High Fault Threshold Register (CJHF)

Write a temperature limit value to this register. When the measured cold-junction temperature is greater than this value, the CJ High fault status bit will be set and (if not masked) the F̅A̅U̅L̅T̅ output will assert.

Default Value: 7Fh

**Register 03h/83h: Cold-Junction High Fault Threshold Register (CJHF) bit map** ([table-17-p22-register.csv](max31856/tables/table-17-p22-register.csv), p. 22)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 03h/83h | CJHF7 | CJHF6 | CJHF5 | CJHF4 | CJHF3 | CJHF2 | CJHF1 | CJHF0 |
|  | Sign | 2^{6} | 2^{5} | 2^{4} | 2^{3} | 2^{2} | 2^{1} | 2^{0} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 04h/84h: Cold-Junction Low Fault Threshold Register (CJLF)

Write a temperature limit value to this register. When the measured cold-junction temperature is less than this value, the CJ Low fault status bit will be set and (if not masked) the F̅A̅U̅L̅T̅ output will assert.

Default Value: C0h

**Register 04h/84h: Cold-Junction Low Fault Threshold Register (CJLF) bit map** ([table-18-p22-register.csv](max31856/tables/table-18-p22-register.csv), p. 22)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 04h/84h | CJLF7 | CJLF6 | CJLF5 | CJLF4 | CJLF3 | CJLF2 | CJLF1 | CJLF0 |
|  | Sign | 2^{6} | 2^{5} | 2^{4} | 2^{3} | 2^{2} | 2^{1} | 2^{0} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 05h/85h: Linearized Temperature High Fault Threshold Register, MSB (LTHFTH)

Write the MSB of the two-byte temperature limit value to this register. When the linearized thermocouple temperature is greater than the two-byte (05h and 06h) limit value, the TC High fault status bit will be set and (if not masked) the F̅A̅U̅L̅T̅ output will assert.

Default Value: 7Fh

**Register 05h/85h: Linearized Temperature High Fault Threshold Register, MSB (LTHFTH) bit map** ([table-19-p22-register.csv](max31856/tables/table-19-p22-register.csv), p. 22)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 05h/85h | LTHFTH7 | LTHFTH6 | LTHFTH5 | LTHFTH4 | LTHFTH3 | LTHFTH2 | LTHFTH1 | LTHFTH0 |
|  | Sign | 2^{10} | 2^{9} | 2^{8} | 2^{7} | 2^{6} | 2^{5} | 2^{4} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 06h/86h: Linearized Temperature High Fault Threshold Register, LSB (LTHFTL)

Write the LSB of the two-byte temperature limit value to this register. When the linearized thermocouple temperature is greater than the two-byte (05h and 06h) limit value, the TC High fault status bit will be set and (if not masked) the F̅A̅U̅L̅T̅ output will assert.

Default Value: FFh

**Register 06h/86h: Linearized Temperature High Fault Threshold Register, LSB (LTHFTL) bit map** ([table-20-p22-register.csv](max31856/tables/table-20-p22-register.csv), p. 22)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 06h/86h | LTHFTL7 | LTHFTL6 | LTHFTL5 | LTHFTL4 | LTHFTL3 | LTHFTL2 | LTHFTL1 | LTHFTL0 |
|  | 2^{3} | 2^{2} | 2^{1} | 2^{0} | 2^{-1} | 2^{-2} | 2^{-3} | 2^{-4} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 07h/87h: Linearized Temperature Low Fault Threshold Register, MSB (LTLFTH)

Write the MSB of the two-byte temperature limit value to this register. When the linearized thermocouple temperature is less than the two-byte (07h and 08h) limit value, the TC Low fault status bit will be set and (if not masked) the F̅A̅U̅L̅T̅ output will assert.

Default Value: 80h Bit 0

**Register 07h/87h: Linearized Temperature Low Fault Threshold Register, MSB (LTLFTH) bit map** ([table-21-p23-register.csv](max31856/tables/table-21-p23-register.csv), p. 23)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 07h/87h | LTLFTH7 | LTLFTH6 | LTLFTH5 | LTLFTH4 | LTLFTH3 | LTLFTH2 | LTLFTH1 | LTLFTH0 |
|  | Sign | 2^{10} | 2^{9} | 2^{8} | 2^{7} | 2^{6} | 2^{5} | 2^{4} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 08h/88h: Linearized Temperature Low Fault Threshold Register, LSB (LTLFTL)

Write the LSB of the two-byte temperature limit value to this register. When the linearized thermocouple temperature is less than the two-byte (07h and 08h) limit value, the TC Low fault status bit will be set and (if not masked) the F̅A̅U̅L̅T̅ output will assert.

Default Value: 00h

**Register 08h/88h: Linearized Temperature Low Fault Threshold Register, LSB (LTLFTL) bit map** ([table-22-p23-register.csv](max31856/tables/table-22-p23-register.csv), p. 23)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 08h/88h | LTLFTL7 | LTLFTL6 | LTLFTL5 | LTLFTL4 | LTLFTL3 | LTLFTL2 | LTLFTL1 | LTLFTL0 |
|  | 2^{3} | 2^{2} | 2^{1} | 2^{0} | 2^{-1} | 2^{-2} | 2^{-3} | 2^{-4} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 09h/89h: Cold-Junction Temperature Offset Register (CJTO)

When the cold-junction temperature sensor is enabled, this register allows an offset temperature to be applied to the measured value. See the Cold-Junction Temperature Sensing section of this data sheet for additional information.

Default Value: 00h

**Register 09h/89h: Cold-Junction Temperature Offset Register (CJTO) bit map** ([table-23-p23-register.csv](max31856/tables/table-23-p23-register.csv), p. 23)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 09h/89h | CJTO7 | CJTO6 | CJTO5 | CJTO4 | CJTO3 | CJTO2 | CJTO1 | CJTO0 |
|  | Sign | 2^{2} | 2^{1} | 2^{0} | 2^{-1} | 2^{-2} | 2^{-3} | 2^{-4} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 0Ah/8Ah: Cold-Junction Temperature Register, MSB (CJTH)

This register contains the MSB of the two-byte (0Ah and 0Bh) value used for cold-junction compensation of the thermocouple measurement. When the cold-junction temperature sensor is enabled, this register is read-only and contains the MSB of the measured cold-junction temperature plus the value in the Cold-Junction Offset register. Also when the cold-junction temperature sensor is enabled, a read of this register will reset the D̅R̅D̅Y̅ pin high. When the cold-junction temperature sensor is disabled, this register becomes a read-write register that contains the MSB of the most recent cold-junction conversion result until a new value is written into it. This allows writing the results from an external temperature sensor, if desired. The maximum contained in the two cold-junction temperature bytes is clamped at 128°C and the minimum is clamped at -64°C.

Default Value: 00h

**Register 0Ah/8Ah: Cold-Junction Temperature Register, MSB (CJTH) bit map** ([table-24-p24-register.csv](max31856/tables/table-24-p24-register.csv), p. 24)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R/W | R/W |
|---|---|---|---|---|---|---|---|---|
| 0Ah/8Ah | CJTH7 | CJTH6 | CJTH5 | CJTH4 | CJTH3 | CJTH2 | CJTH1 | CJTH0 |
|  | Sign | 2^{6} | 2^{5} | 2^{4} | 2^{3} | 2^{2} | 2^{1} | 2^{0} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 0Bh/8Bh: Cold-Junction Temperature Register, LSB (CJTL)

This register contains LSB of the two-byte (0Ah and 0Bh) value used for cold-junction compensation of the thermocouple measurement. When the cold-junction temperature sensor is enabled, this register is read-only and contains the LSB of the measured cold-junction temperature plus the value in the Cold-Junction Offset register. Also when the cold-junction temperature sensor is enabled, a read of this register will reset the D̅R̅D̅Y̅ pin high. When the cold-junction temperature sensor is disabled, this register becomes a read-write register that contains the LSB of the most recent cold-junction conversion result until a new value is written into it.

Default Value: 00h

**Register 0Bh/8Bh: Cold-Junction Temperature Register, LSB (CJTL) bit map** ([table-25-p24-register.csv](max31856/tables/table-25-p24-register.csv), p. 24)

| MEMORY ACCESS | R/W | R/W | R/W | R/W | R/W | R/W | R | R |
|---|---|---|---|---|---|---|---|---|
| 0Bh/8Bh | CJTL7 | CJTL6 | CJTL5 | CJTL4 | CJTL3 | CJTL2 | CJTL1 | CJTL0 |
|  | 2^{-1} | 2^{-2} | 2^{-3} | 2^{-4} | 2^{-5} | 2^{-6} | 0 | 0 |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 0Ch: Linearized TC Temperature, Byte 2 (LTCBH)

This  is  the  high  byte  of  the  19-bit  register  that  contains  the  linearized  and  cold-junction-compensated  thermocouple temperature value.

Default Value: 00h

**Register 0Ch: Linearized TC Temperature, Byte 2 (LTCBH) bit map** ([table-26-p24-register.csv](max31856/tables/table-26-p24-register.csv), p. 24)

| MEMORY ACCESS | R | R | R | R | R | R | R | R |
|---|---|---|---|---|---|---|---|---|
| 0Ch | LTCBH7 | LTCBH6 | LTCBH5 | LTCBH4 | LTCBH3 | LTCBH2 | LTCBH1 | LTCBH0 |
|  | Sign | 2^{10} | 2^{9} | 2^{8} | 2^{7} | 2^{6} | 2^{5} | 2^{4} |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

### Register 0Dh: Linearized TC Temperature, Byte 1 (LTCBM)

This is the middle byte of the 19-bit register that contains the linearized and cold-junction-compensated thermocouple temperature value.

Default Value: 00h

 **Register 0Dh: Linearized TC Temperature, Byte 1 (LTCBM) bit map** — [table-27-p25-register.csv](max31856/tables/table-27-p25-register.csv), p. 25

| MEMORY ACCESS   | R      | R      | R      | R      | R      | R      | R      | R      |
|-----------------|--------|--------|--------|--------|--------|--------|--------|--------|
| 0Dh             | LTCBM7 | LTCBM6 | LTCBM5 | LTCBM4 | LTCBM3 | LTCBM2 | LTCBM1 | LTCBM0 |
|                 | 2^{3}    | 2^{2}    | 2^{1}    | 2^{0}    | 2^{-1}   | 2^{-2}   | 2^{-3}   | 2^{-4}   |

### Register 0Eh: Linearized TC Temperature, Byte 0 (LTCBL)

This is the low byte of the 19-bit register that contains the linearized and cold-junction-compensated thermocouple temperature value.

Default Value: 00h

**Register 0Eh: Linearized TC Temperature, Byte 0 (LTCBL) bit map** ([table-28-p25-register.csv](max31856/tables/table-28-p25-register.csv), p. 25)

| MEMORY ACCESS | Bit 7 |  |  |  |  |  |  | Bit 0 |
|---|---|---|---|---|---|---|---|---|
|  | R | R | R | R | R | R | R | R |
| 0Eh | LTCBL7 | LTCBL6 | LTCBL5 | LTCBL4 | LTCBL3 | LTCBL2 | LTCBL1 | LTCBL0 |
|  | 2^{-5} | 2^{-6} | 2^{-7} | X | X | X | X | X |

### Register 0Fh: Fault Status Register (SR)

The  Fault  Status  Register  contains  eight  bits  that  indicate  the  fault  conditions  (Thermocouple  Out-of-Range,  Cold Junction Out-of-Range, Cold Junction High, Cold Junction Low, Thermocouple High Temperature, Thermocouple Low Temperature, Over-Under Voltage, or Open Thermocouple) that have been detected.

Default Value: 00h Bit 0

**Register 0Fh: Fault Status Register (SR) bit map** ([table-29-p25-register.csv](max31856/tables/table-29-p25-register.csv), p. 25)

| MEMORY ACCESS | R | R | R | R | R | R | R | R |
|---|---|---|---|---|---|---|---|---|
| 0Fh | CJ Range | TC Range | CJHIGH | CJLOW | TCHIGH | TCLOW | OVUV | OPEN |
|  | Bit 7 |  |  |  |  |  |  | Bit 0 |

- Note: When the MAX31856 is set to operate in "comparator" fault mode (set with bit 2 of Configuration 0 register (00h)), the fault status bits simply reflect the state of any faults by asserting when the fault condition is true, and deasserting when the fault condition is no longer true.

- Note: When in 'interrupt' fault mode, the fault status bits assert when a fault condition is true. The bits remain asserted until a 1 is written to the Fault Status Clear bit. This deasserts the fault bits until a new fault is detected (note that this may occur immediately if the fault condition is still in place).

### Register 0Fh: Fault Status Register (SR) (continued)

**Register 0Fh: Fault Status Register (SR) (continued) — Bit Descriptions** ([table-30-p26-register.csv](max31856/tables/table-30-p26-register.csv), p. 26)

| BIT | NAME | DESCRIPTION |
|---|---|---|
| 7 | CJ Range | Cold Junction Out-of-Range 0 = The Cold-Junction temperature is within the normal operating range (-55°C to +125°C for types E, J, K, N, and T; -50°C to +125°C for types R and S; 0 to 125°C for type B). 1 = The Cold-Junction temperature is outside of the normal operating range. |
| 6 | TC Range | Thermocouple Out-of-Range 0 = The Thermocouple Hot Junction temperature is within the normal operating range (see Table 1). 1 = The Thermocouple Hot Junction temperature is outside of the normal operating range. Note: The TC Range bit should be ignored in voltage mode. |
| 5 | CJHIGH | Cold-Junction High Fault 0 = The Cold-Junction temperature is at or lower than the cold-junction temperature high threshold (default). 1 = The Cold-Junction temperature is higher than the cold-junction temperature high threshold. The F̅A̅U̅L̅T̅ output is asserted unless masked. |
| 4 | CJLOW | Cold-Junction Low Fault 0 = The Cold-Junction temperature is at or higher than the cold-junction temperature low threshold (default). 1 = The Cold-Junction temperature is lower than the cold-junction temperature low threshold. The F̅A̅U̅L̅T̅ output is asserted unless masked. |
| 3 | TCHIGH | Thermocouple Temperature High Fault 0 = The Thermocouple Temperature is at or lower than the thermocouple temperature high threshold (default). 1 = The Thermocouple Temperature is higher than the thermocouple temperature high threshold. The F̅A̅U̅L̅T̅ output is asserted unless masked. |
| 2 | TCLOW | Thermocouple Temperature Low Fault 0 = Thermocouple temperature is at or higher than the thermocouple temperature low threshold (default). 1 = Thermocouple temperature is lower than the thermocouple temperature low threshold. The F̅A̅U̅L̅T̅ output is asserted unless masked. |
| 1 | OVUV | Overvoltage or Undervoltage Input Fault 0 = The input voltage is positive and less than V_{DD} (default). 1 = The input voltage is negative or greater than V_{DD}. The F̅A̅U̅L̅T̅ output is asserted unless masked. Note: The presence of the OVUV fault will suspend conversions and the ability of the MAX31856 to detect other faults (or clear faults when in comparator mode) until the fault is no longer present. |
| 0 | OPEN | Thermocouple Open-Circuit Fault 0 = No open circuit or broken thermocouple wires are detected (default) 1 = An open circuit such as broken thermocouple wires has been detected. The F̅A̅U̅L̅T̅ output is asserted unless masked. |

### Applications Information

### Thermocouple Temperature Sensing Guidelines

Follow these guidelines to get the best results when sensing temperature. The Typical Application Circuit shows a basic  MAX31856 schematic. Connect the thermocouple wires to inputs T+ and T-; be sure that the wires are connected to the correct input as shown in Figure 8. Connect the BIAS output to T-. This biases the thermocouple within the common-mode range of the inputs.

### Noise Considerations

Because of the small signal levels involved, thermocouple temperature  measurement  is  susceptible  to  power-supply-coupled noise. The effects of power-supply noise can be minimized by placing 0.1µF ceramic bypass capacitors close to the V_{DD} pins and to GND.

The input  amplifier  is  a  low-noise  amplifier  designed  to enable high-precision input sensing. Keep the thermocouple and connecting wires away from electrical noise sources. It is strongly recommended to add a 100nF ceramic surface-mount differential capacitor, placed across the T+ and T- pins, to filter noise on the thermocouple lines. In environments with high noise levels, especially significant RF fields, a 100nF capacitor between T+ and T- should be supplemented with a 10nF capacitor between T+ and GND, and another 10nF capacitor between T- and GND. These values may need to be modified depending on the nature  of  the  noise  pickup.  Other  techniques,  such  as adding series resistance and shielding the thermocouple wires  and  circuit  board,  may  also  be  necessary  in  the presence  of  larger  noise  sources.  Figure  8  shows  the typical application circuit with input capacitors and input resistors added. 

### Input Protection

The ±45V input protection  circuitry  prevents  damage  to the  IC  caused  by  overvoltage  conditions  at  T+,  T-,  or BIAS. If larger input faults are possible, external protection should be added. Resistors in series with T+, T-, and BIAS  can  increase  the  acceptable  fault  voltages.  For example, adding 2kΩ in series with these inputs allows an  additional  ±40V  of  overdrive  before  the  20mA  input current  limit  is  reached.  Note,  however,  that  if  the  input has 45V across it and 20mA flowing into it, the power dissipation will be 900mW due to the overdrive at that input. Overdriving  other  inputs  at  the  same  time  will  further increase  the  power  dissipation. Always  ensure  that  if  a continuous overdrive voltage greater than ±45V is expected, any current-limiting resistors are large enough to keep total power dissipation well under the IC's absolute maximum power dissipation. Note also that added resistance in  series  with T+ and T- can increase offset voltage, as mentioned in the Effect of Series Resistance section.

![Figure 8. Typical Connection to Reduce the Effect of Noise Pickup in the Thermocouple Cable](max31856/figures/max31856-f032.png)

*Figure 8. Typical Connection to Reduce the Effect of Noise Pickup in the Thermocouple Cable*

![Figure 9. When Thermocouple Inputs May Be Exposed to Fault Voltages Greater than ±45V, Resistors Can be Added to Limit Current into the MAX31856.](max31856/figures/max31856-f033.png)

*Figure 9. When Thermocouple Inputs May Be Exposed to Fault Voltages Greater than ±45V, Resistors Can be Added to Limit Current into the MAX31856.*

### Effect of Series Resistance

Bias and leakage current at the thermocouple inputs will flow  through  input  resistors  and  cable  resistance,  generating  input  offset  voltage.  For  the  circuits  in  Figure  8 and Figure 9, assuming that the thermocouple's source resistance  is  negligible,  the  offset  voltage  due  to  series resistance will be:

![Equation as printed (p. 28)](max31856/figures/max31856-eq01.png)

I_{B} x ∆R_{S} + ∆I_{B} x R_{S} 

where:

- R_{S} is the series resistance between each input and the bias point
- ∆R_{S} is the difference between the two R_{S} values. This will generally be equal to the tolerance of any discrete series resistors plus any cable resistance.
- I_{B} is the input bias and leakage current
- ∆I_{B} is the differential input bias and leakage current

As an example, assume that the circuit in Figure 8 will be used up to a temperature of 85°C, the mismatch between the  100Ω  input  resistors  is  1Ω,  and  the  external  cable resistance is 50Ω. This yields a worst-case offset voltage due to the external resistances of:

65nA x (50Ω + 1Ω) + 4nA x 100Ω = 3.7μV

*[Equation not captured in extraction — see the PDF.]*

To minimize the effect of input resistance on accuracy:

- Minimize the values of any external resistors
- When the cable resistance is very low, match the values of the external resistors as closely as possible.
- If the cable resistance is known, increase the value of the resistor connected to T- by the value of the cable resistance. This will minimize the total mismatch between the two inputs.

If the cable resistance is excessive, consider using largergauge thermocouple wire.

### MAX31856 Location

Because the MAX31856 includes an internal cold-junction temperature sensor, place it in a location whose temperature is as close as possible to that of the cold junction. If the thermocouple leads are directly soldered to the PCB, the  MAX31856  should  be  as  close  as  possible  to  the thermocouple  lead  connections  and  thermal  gradients between the IC and the thermocouple connections should be  minimized.  If  the  thermocouple  leads  terminate  in  a connector, mount the IC as close as possible to the connector, and again minimize thermal gradients between the connector and the IC.

### Using 'Unsupported' Thermocouple Types

To  use  a  thermocouple  type  other  than  B,  E,  J,  K,  N, R,  S,  or  T,  select  one  of  the  voltage  mode  options  in Configuration 1. Selecting 'Gain = 8' results in a full-scale input  voltage  range  of  ±78.125mV.  'Gain  =  32'  results in  a  full-scale  input  voltage  range  of  ±19.531mV.  See

### Ordering Information

**Ordering Information** ([table-31-p29-data.csv](max31856/tables/table-31-p29-data.csv), p. 29)

| PART | TEMP RANGE | PIN-PACKAGE |
|---|---|---|
| MAX31856MUD+ | -55°C to +125°C | 14 TSSOP |
| MAX31856MUD+T | -55°C to +125°C | 14 TSSOP |

+Denotes a lead(Pb)-free/RoHS-compliant package.  

T = Tape and reel.

the  transfer  functions  from  the  Configuration  1  Register table. When voltage mode is selected, no linearization is performed on the conversion data. Use the voltage data and the cold-junction temperature to calculate the thermocouple's hot-junction temperature.

### Package Information

For  the  latest  package  outline  information  and  land  patterns (footprints), go to www.maximintegrated.com/packages . Note that a '+', '#', or '-' in the package code indicates RoHS status only. Package drawings may show a different suffix character, but the drawing pertains to the package regardless of RoHS status.

**Package Information** ([table-32-p29-data.csv](max31856/tables/table-32-p29-data.csv), p. 29)

| PACKAGE TYPE | PACKAGE CODE | OUTLINE NO. | LAND PATTERN NO. |
|---|---|---|---|
| 14 TSSOP | U14+2 | 21-0066 | 90-0113 |

### Revision History

**Revision History** ([table-33-p30-data.csv](max31856/tables/table-33-p30-data.csv), p. 30)

| REVISION NUMBER | REVISION DATE | DESCRIPTION | PAGES CHANGED |
|---|---|---|---|
| 0 | 2/15 | Initial release | — |

For pricing, delivery, and ordering information, please contact Maxim Direct at 1-888-629-4642, or visit Maxim Integrated's website at www.maximintegrated.com.

Maxim Integrated cannot assume responsibility for use of any circuitry other than circuitry entirely embodied in a Maxim Integrated product. No circuit patent licenses are implied. Maxim Integrated reserves the right to change the circuitry and specifications without notice at any time. The parametric values (min and max limits) shown in the Electrical Characteristics table are guaranteed. Other parametric values quoted in this data sheet are provided for guidance.